A Practical Guide to AFM Probes and Sensors
Paul West, Pacific Nanotechnology, Tustin, California
Most atomic force microscopes (AFMs) use a probe mounted at the end
of a cantilever for measuring images. Probe/cantilever combinations
are available in various sizes and shapes, with and without coatings,
and choosing the right one for a particular application is straightforward.
Using the wrong sensor can result in distorted or misleading images.
Sensor
selection involves choosing a scanning mode (contact or vibration) and
defining the application (sample type, image resolution required and
the material characteristics of interest). Probe tips are matched to
the topography of the sample and the degree of resolution required and
range from rounded, to sharp, to ultra-sharp and high-aspect ratio tips,
including carbon nanotubes.
An AFM
can image many characteristics of a surface, including physical topography,
magnetization, electrostatic force and tribological properties such
as friction, lubrication and wear. AFMs can also sense and image material
compositions in some cases and quantitatively measure surface roughness
and the dimensions of surface features.
AFM Fundamentals
An atomic
force microscope scans a probe mounted on a cantilever (Figure 1)
over a surface; the forces between the probe tip and the sample surface
are monitored, and an image of the surface is generated. The most common
type of force sensor utilizes the relationship between the resulting
motion of the cantilever and the applied force. The motion is measured
with the "light lever" method in which a photo-detector measures
the light reflected from the back side of the cantilever, and the output
of the photo-detector is proportional to the motion of the cantilever.
Motions as small as 1 nm are routinely measured with the light lever
method.

|
|
|
Figure
1: An AFM sensor includes a cantilever with a mounted probe.
|
This combination of probe and cantilever can be monolithic - manufactured
by MEMS techniques from single-crystal silicon - or the probe and cantilever
can be of different materials. An example of the latter is a carbon
nanotube (CNT) tip mounted on a silicon or silicon nitride cantilever.
This probe tip is used when extremely high resolution is required.
Generally,
the sharper the probe tip, the better the image will be (Figure 2).
However, images are actually a combination of probe tip shape and sample
topography and the best probe tip shape is not always a sharp point
(Figure 3).When imaging rectangular types of features such as
those found in many semiconductor chip applications, a more rectangular
shape is required. In this case the quality of tip shape is just as
important, but the focus is on edge shape rather than point sharpness.
An inappropriate probe tip shape or a damaged tip will not follow the
surface contour and will produce a distorted image.

|
|
|
Figure
2: Generally, a sharper probe tip will produce a higher resolution
image.
|

|
|
|
Figure
3: The probe tip shape should be appropriate to the surface
topography and a sharp point is not always best. Left: steep sidewalls
as on a semiconductor chip require a rectangular probe tip. Right:
A relatively smooth surface with shallow topography requires the
more common, sharp probe tip.
|
The strongest
forces between a particular probe and the surface are mechanical; however,
other forces can provide information about surface properties other
than topography. Obtaining these kinds of information is a combination
of AFM operation protocols and sensor characteristics. Images can show
hardness, electrical and magnetic properties, tribology, and other material properties. It is also possible to "sense" the materials at
a sample's surface. If a surface is perfectly flat, but has an interface
between two different materials, the image can show the change in material
composition.
A more
extensive tutorial on atomic force microscopy can be found at www.pacificnanotech.com.
Physical
Characteristics
Cantilevers
are approximately 125 µm to 450 µm long - slightly larger
than the width of a human hair. The probe itself is 1000 times smaller.
Typical probe tips have radii of less than 10 nm (1 nm = 1 billionth
of a meter; a carbon atom is about 0.25 nm in diameter.). In addition
to the probe tip diameter, the aspect ratio, which is the ratio of probe
tip length to its diameter, is important.
Over the
years, industry-standard products have emerged for sensors commonly
used in atomic force microscopes. The silicon monolithic design is most
common (Figure 4). The cantilever and probe are supported by
a 1.6 mm x 3.4 mm single-crystal silicon holder (called the substrate
or chip). The chip permits easy replacement of the tip without a major
readjustment of the detection system.

|
|
|
Figure
4: The monolithic sensor structure is the most commonly used
sensor in AFMs.
|
The silicon
is highly doped for electrical conductivity and has a typical resistivity
of 0.01 to 0.025 ohm-cm. The structure has no intrinsic stress and,
as a result, the cantilever is absolutely straight. Silicon is chemically
inert for most applications. The cantilever itself is rectangular and
usually has a tapered free end for visibility during positioning. The
cantilever has a trapezoidal cross section with a wide side to reflect
the laser beam that detects its motion. Cantilever width is usually
given as the average (mean) of the two sides of the trapezoid. Sensors
with a reflective (or reflex) coating have a layer of aluminum, typically
about 30 nm thick, on the backside of the cantilever to enhance their
reflectivity from 2 to 2.5 times. However, corrosive environments can
attack the aluminum coating and uncoated sensors are available for these
applications.
A typical
probe tip height is 10-15 µm with a radius of curvature of less
than 10 nm at the probe tip end. The macroscopic half-cone angles are
20-25° viewed along the cantilever axis and 25-30° seen from
the side, and virtually zero when viewed from the probe tip end (Fig.
5-1). Since the cantilever is flexed in use, the probe is mounted at
a slight angle (Fig. 5-3). Probe tips can be sharper, with a radius
less than 5 nm and a half cone angle less than 10° at the final
100 nms. A sharper probe tip gives enhanced resolution of micro-roughness
and nanostructures.
High-aspect-ratio
probe tips are available for measurements on samples with sidewall angles
approaching 90° such as deep, narrow trenches on semiconductor structures
(Fig. 5-2). They have an overall height of 10-15 µm to allow measurements
in deep crevices and corrugations. The last few µm have a symmetrical
high aspect ratio with a radius of about 10-15 nm. The high aspect ratio
portion of the probe tip is 1.5 µm to 2.0 µm long. The length/diameter
aspect ratios can be as great as 12:1 resulting in half cone angles
as small as 2.8°.
To compensate
for the tilt angle of the cantilever when mounted on the AFM head, the
high-aspect-ratio probe can be "tilt compensated," that is,
mounted so that the high aspect portion of the probe tip is tilted about
13 ° with respect to their center axis (Figure 6). This allows symmetrical
imaging and accurate slope measurement of near-vertical sidewalls.

|
|
|
Figure
5: These are scanning electron microscope (SEM) photos of
typical silicon probes and probe tips. (1). Standard pyramidal
probe tip. (2). High-aspect-ratio probe tip. (3). Probe at end
of cantilever. (4). View of a standard probe from the end of the
cantilever.
|

|
|
|
Figure
6: Tilt-compensated high-aspect-ratio probe tips deal with
the tilt angle of the cantilever when mounted on the AFM head.
|
Generally,
smaller (sharper) probe tips and probe tips with higher aspect ratios
are more expensive.
Image
Resolution
In-plane
image resolution depends on the relation of the probe tip geometry -
the sharpness, shape, and aspect ratio - to the topography of the surface.
The bottom of the probe tip must be able to follow the surface contours
for an accurate image to be produced. Figure 7 shows a perovskite surface
scanned with a 5 nm diameter probe tip and a different, but similar,
region on the same sample scanned with a 40nm diameter probe tip.

|
|
|
Figure
7: AFM images of perovskite surface. A probe tip with a 40nm
diameter was used to measure the left image and a probe tip with
5 nm diameter was used to measure the probe at the right side.
|
Scanning
Modes
The sensor
must be matched to the scanning mode and there are two principal scanning
methods: contact (or deflection) mode and non-contact (or vibration)
mode (Figure 8). In contact mode, the probe is scanned over the
surface with a fixed cantilever deflection. The force on the surface
is directly proportional to the deflection and is extremely small, often
less than a nano-newton. The probe tip is minimally touching the surface.
In the vibration mode (also called close-contact, or intermittent contact)
the cantilever oscillates at its resonant frequency during scanning.
The resonance changes as the probe tip approaches the surface and the
vibration amplitude and phase are measured. Although the non-contact
mode is often called "tapping", it is important that the probe
tip not "tap" the sample or it may be broken or the sample
damaged.

|
|
|
Figure
8: The two common scanning modes are contact mode (left) and
vibration or non-contact mode (right).
|
Probe tips
used in contact mode are more subject to wear-out, and wear-resistant
diamond coatings are available. All sensors are fragile, but in general
their fragility increases with the sharpness and aspect ratio of the
probe tip.
Vibration
mode is used with very soft samples. Also, in combination with very
sharp probe tips, vibration mode is preferred to obtain high image resolutions
(<10nm).
In vibration mode, the magnitude of amplitude damping and the amount
of phase change of the cantilever depends on the surface chemical composition
and the physical properties of the surface. On an inhomogeneous sample,
contrast can be observed between regions of varying chemical composition.
This technique has many names including phase mode, phase detection
and force modulated microscopy.
Coatings
Probes
and cantilevers are available with several coatings for various applications.
The most commonly used coating is the reflex coating for improved image
quality, but it can't be used in a corrosive environment. Other coatings
are required to image some sample characteristics other than topography
and/or deal with sample hardness. The common coatings encountered in
choosing the best AFM sensor are described below:
Reflex Coatings for Image Quality - A 30-nm thick, stress-free aluminum
reflex (reflective) coating on the detector side of the cantilever can
improve the reflectivity of the laser beam by up to 2-½ times
and increase the signal/noise of the AFM. The coating prevents light
interference, which degrades the reflected signal. However, reflex coatings
cannot be used in corrosive environments that react with the aluminum.
An uncoated cantilever is nearly inert.
PrIr5 Coating
for Enhanced Electrical Contact and Conductivity - The PrIr5 coating
is an approximately 23 nm thick double layer of chromium and platinum
iridium5 on both sides of the cantilever. The probe-side coating enhances
the conductivity of the probe and allows electrical contact. The detector-side
coating also enhances the reflectivity of the laser beam by a factor
of 2 and is corrosion resistant. The PtIr5 layer is stress compensated
and wear resistant; cantilever bending due to stress is typically less
than 2°. Electrostatic Force Microscopy (EFM) applications would
always use this coating.
Diamond
Coating for Wear Resistance - For applications that require hard contact
between the probe tip and sample, a 100 nm thick, diamond probe-side
coating provides extremely high wear resistance. Applications include
friction measurements, the measurement of elastic properties of samples
that are not too soft, and wear measurements.
Diamond
Coating for Conductivity - Special conductive diamond-coated probes
are available when there will be hard contact between the probe tip
and sample and electrical conductance is also required. Examples of
typical applications include Scanning Spreading Resistance Microscopy
(SSRM), Tunneling AFM, and Scanning Capacitance Microscopy (SCM).
Hard Magnetic
Coating for Magnetic Force Microscopy (MFM) - Investigating the magnetic
properties of surfaces on the nanometer scale requires a hard magnetic
coating on the probe. Typically, about 40 nm of cobalt alloy provides
a permanent magnetic field at the probe tip. An example is the measurement
of magnetic domains on magnetic recording media. The probe must be magnetized
by means of an external strong magnet. Soft magnetic samples may be
influenced by the probe tip magnetization.
Other
Special Sensors
Some applications
require that the cantilever have special characteristics. For example,
Lateral Force Microscopy (LFM) sensors have an extremely soft, thin
cantilever to provide high sensitivity to lateral or friction forces.
Force Modulation Mode (FMM) sensors have force constants tailored to
their application.
Sensors
for special and research applications are available including cantilevers
without probes, rounded probe tips having radii from 50 nm to 500 nm,
and plateau probe tips with a flattened end from 1 µm to 8 µm
wide.
Force
Constant and Resonant Frequency
Cantilevers
are characterized by their force constant and resonant frequency and
are matched to the operating mode and AFM instrument. The longer the
cantilever, the lower the resonance frequency. Typical characteristics
and operating modes are listed in Table I.
| Table
I. Typical Cantilever Sensor Characteristics |
| Mode
|
Force
Constant
|
Resonant
Frequency
|
| Contact
Mode |
0.2
N/m
1.6 N/m
|
13-14
kHz
27 kHz
|
| Non-Contact |
42
N/m
48 N/m
15 N/m
27 N/m
|
330
kHz
190 kHz
130 kHz
|
| Force
Modulation |
2.8
N/m
|
75
kHz
|
| Lateral
Force (Torsion) |
0.2
N/m
|
25
kHz
|
| Electrostatic
or Magnetic Force |
2.8
N/m
|
75
kHz
|
|
Some AFM
microscopes require a minimum cantilever length (> 125 µm)
and/or do not accept high frequencies. Sensors can be obtained in a
choice of lengths and resonant frequencies.
Sensor
Management in the Microscope
Sensors
can be purchased in small packages, typically 10 to 50 per package or
as silicon wafers containing several hundred cantilevers with integral
probes. Usually, yield from the wafer is guaranteed to be a fixed number
or 80-90% of the total. Purchasing unmounted sensors in wafer form saves
money but adds labor in terms of handling and mounting.
AFM sensors
are very small and hard to pick up and put in the microscope. The sensor
chip on which the cantilever is mounted typically measures 1 x 2-4 mm
and can be easily dropped. To simplify mounting and prevent damage to
sensors, Pacific Nanotechnology offers sensors mounted on a larger metal
substrate that is held in the microscope by magnets. If it is necessary
to use un-mounted sensors, the Pacific Nanotechnology mounting tool
may be used for mounting AFM cantilevers to a clip holder. (Figure 9)
To operate the probe fixture tool, the probe clip holder is placed into
the fixture. Then the probe is placed onto the fixture; finally the
probe can easily slide into the probe clip holder.
Sensor
lifetime is an important issue in any AFM application, and how long
a probe tip lasts is related to the skill of the microscope operator.
Lifetimes increase with careful probe tip approach to the surface, by
preventing the probe tip from tapping the surface with excess force
in vibration mode and by scanning at speeds appropriate to the sensor's
physical characteristics (i.e. not too fast). Probe tips can last a
long time with careful use. More expensive, special-application sensors
can be removed and reinserted for use only when required by the application.
Some AFMs
are used for one type of application in dedicated research environments
or with only one or two users doing routine examination of similar samples.
Other AFMs see many users and perform many different applications in
a single day. Most fall somewhere in between.
The challenge
for instruments used for a wide variety of surfaces is to minimize sensor
changes and sensor inventory. It may be feasible, now that AFMs are
becoming more affordable, to have more than more instrument and dedicate
one to advanced applications and one to routine imaging.

|
|
|
Figure
9: This probe exchange tool, opened with probe clip holder
sitting in front is an example of a tip mounting tool provided
by Pacific Nanotechnology.
|
Flatness
standards, height standards, and several standards with 100, 200, or
300 pitch for precise x-y calibration of the microscope's scanning mechanism
are readily available.
Glossary
of Terms Relating to AFM Sensors
AFM
- Atomic Force Microscopy or Atomic Force Microscope - the technique
or microscope used to image topography and many other material characteristics
of a surface on the nanometer to micrometer scale.
Sensor
- The combination of a probe/probe tip, cantilever, and substrate (chip)
used to measure the forces from which an AFM image is developed.
Probe
- The structure, mounted on the end of an AFM cantilever, that interacts
with the sample surface being imaged. The term AFM probe refers to the
entire structure including the probe tip.
Probe
Tip - The outer end of the AFM probe that interacts directly with
the sample surface during imaging. Its shape is the most critical aspect
in choosing the best sensor for a particular application. Sometimes
also referred to as a sensor tip.
Cantilever
- The part of an AFM Sensor on which the probe is mounted. The motion
of the cantilever is the metric used to measure the forces between the
probe tip and the sample surface.
Substrate
or Chip - Terms used interchangeably to refer to the support on
which a cantilever and probe and probe tip are mounted.